12.13 Technical Information
Electromagnetic Compatibility (EMC) Information
Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC). Radio frequency (RF) communications equipment can affect
devices such as Rythmic™ Evolution pump. As such, the pump should not be adjacent
to these type of equipment. If it is not practical, then observe the pump to make sure it is
operating properly after installation.
Guidance and manufacturer’s declaration: electromagnetic emissions
Important
The use of non-recommended accessories may result in increased EMC emissions
or decreased EMC immunity of the Rythmic™ pump. Refer to Chapter 10, paragraph
10.7 "Accessories"
Emissions test Compliance Electromagnetic environment/guidance
RF emissions
CISPR 11
Group 1
The Rythmic™ Evolution pump uses RF energy
only for its internal functioning. Therefore its
RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Rythmic™ Evolution pump is suitable for
use in all establishments, including domestic
surroundings.
Harmonic
emissions
IEC 61000-3-2
Class B
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies