6.2 Placement of the pump
All Rythmic™ Organiser pumps should only be
kept in a vertical position while using with a rigid
container which could contain air (e.g. glass
bottles, hard plastic bag). Non vertical operation of
the Rythmic™ Organiser pump when rigid external
medication container is used, could result in under
or non-delivery of medications.
Please refer to Chapter 10, paragraph 10.5 "Using
carrying bag and carrying belt" for details regarding
the use of carrying bags for portable use.
!Warning
Ensure that the spike connector is properly attached to the external medication
container. Spike detachment from the external medication container could result
in under or non-delivery of medication.
When rigid - non collapsible external medication containers (e.g. glass bottles, hard
plastic bag) are used with a Rythmic™ spike set, then the air vent on the spike must be
opened and the container must be suspended from a pole. A closed air vent could result
in under or non-delivery of medication. The air vent is fitted with an anti-bacterial filter.
When rigid – non collapsible external medication containers (e.g. glass bottles, hard
plastic bag) are used in ambulatory infusion (infusion during transportation), ensure that
the external container is always placed vertical.
In Rythmic™ Evolution Organiser pump range, the container must be suspended by
the bag hook placed inside the pump’s transparent cover.
All Rythmic™ Evolution Organiser pumps should be placed in pole, during a non-
ambulatory infusion.
All Rythmic™ Evolution Organiser pumps can be transported using carrying belt, handle
or carrying bag. Ensure that the container inside the Organiser remains always in vertical
position.
Only collapsible containers should be used when Mini Rythmic™ Evolution pump
operates in waist-shoulder carrying bag.
Ensure that the transparent covers are securely closed, to prevent patient access to
medication container.
Ensure that the administration set has been properly installed and has not been
trapped in transparent covers or inside the carrying bag, as this may result in under or
non-delivery of medication.
Ensure that during infusion, the vertical distance between the pump and the injection
site should be one meter height maximum.
In all above cases, failure to follow the instructions may result in under or non-delivery
of medication and air in line may occur, which could result in death or serious injury to the
patient.