4. Actions and codes
Pump has different codes to secure access to certain programming and operations.
Depending on the desired functionality, the appropriate code must be used. The
table below lists the Actions possible and the corresponding code that must be
entered to accomplish them.
Actions available after turning On the pump
After turning on the pump, a code must be entered by the user.
Actions available after initiate infusion
!Warning
We recommend the default codes of the pump to be changed (through the configuration menu)
in order to secure the access to different actions related to each code.
The supervisor of the infusion should decide to whom each code and its related actions
should be distributed. Incorrect pump programming could result in death or serious injury
to the patient.
CODE:
0000
Programming
Code
New bag
code
Resume
Code
Select
mode code
Select delivery mode
x
Program an infusion
x
New bag
x
x
Resume an infusion
x x
Programming
Code
New bag
code
Resume
Code
Select
mode code
Titrate a running protocol
x
Clinician bolus
x
New bag
x
Clear counters
x
Stop infusion, if code
protected stop of infusion is
enable
x x
34 35
Actions
Actions
Default codes
Default codes

Table of Contents