EasyManua.ls Logo

Micrel Rhythmic evolution - Historic

Default Icon
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Loading dose
OFF or from 0.1 ml to 60.0 ml 0.1 ml inc.
0.1- 999.9 mg 0.1 mg inc. or 0.1- 999.9 μg 0.1 μg inc.
Clinician Override Bolus
OFF or from 0.1 ml to 60 ml 0.1 ml inc.
0.1- 999.9 mg 0.1 mg inc. or 0.1- 999.9 μg 0.1 μg inc.
Maximum infusion pressure 2 bar (1500 mm Hg, 29 psi, 200 kPa)
Occlusion alarm (3levels)
Low: 0.8
+0,4
/-0.4 bar (600 mmHg
+300
/-300,
11.6 psi
+5.80
/-5,80, 80 kPa
+40
/-40,)
Middle:1.2
+0,4
/-0.4 bar (900 mmHg
+300
/-300,
17.4 psi
+5.80
/-5.80, 120 kPa
+40
/-40,)
High: 1.6
+0,4
/-0.4 bar (1200 mmHg
+300
/-300,
23.2 psi
+5.80
/-5.80, 160 kPa
+40
/-40)
Coded access 3 levels
12.4 Historic
Infusion monitoring
• Patient Bolus demand
• Patient Bolus delivered
• Total infused in ml and mg or μg
• Volume to be infused
• Hour by hour delivery analysis
• Graphical infusion trend analysis
Event Log
• 6500 time stamped events
• Protocols and protocol changes
• Alarms
• Patient bolus attempt and patient bolus delivered.
• ON OFF, Loading dose, Clinician dose, set change
PC connection
• RS232 output
• Data download
•
9600 bds, 8 bit, no parity, one stop bit, Xon/X off flow control
92 93

Table of Contents