Loading dose
OFF or from 0.1 ml to 60.0 ml 0.1 ml inc.
0.1- 999.9 mg 0.1 mg inc. or 0.1- 999.9 μg 0.1 μg inc.
Clinician Override Bolus
OFF or from 0.1 ml to 60 ml 0.1 ml inc.
0.1- 999.9 mg 0.1 mg inc. or 0.1- 999.9 μg 0.1 μg inc.
Maximum infusion pressure 2 bar (1500 mm Hg, 29 psi, 200 kPa)
Occlusion alarm (3levels)
Low: 0.8
+0,4
/-0.4 bar (600 mmHg
+300
/-300,
11.6 psi
+5.80
/-5,80, 80 kPa
+40
/-40,)
Middle:1.2
+0,4
/-0.4 bar (900 mmHg
+300
/-300,
17.4 psi
+5.80
/-5.80, 120 kPa
+40
/-40,)
High: 1.6
+0,4
/-0.4 bar (1200 mmHg
+300
/-300,
23.2 psi
+5.80
/-5.80, 160 kPa
+40
/-40)
Coded access 3 levels
12.4 Historic
Infusion monitoring
Patient Bolus demand
Patient Bolus delivered
Total infused in ml and mg or μg
Volume to be infused
Hour by hour delivery analysis
Graphical infusion trend analysis
Event Log
6500 time stamped events
Protocols and protocol changes
Alarms
Patient bolus attempt and patient bolus delivered.
ON OFF, Loading dose, Clinician dose, set change
PC connection
RS232 output
Data download
9600 bds, 8 bit, no parity, one stop bit, Xon/X off flow control