9. Configuration
This chapter refers to pump configurations, which allows the pump to be customised
to user and/ or patient specificities and must be manipulated by the clinician only.
!Warning
The supervisor of the infusion must decide whether the Configuration chapter should be
kept in the present IFU.
9.1 Definitions
KVO
A Keep venous Open rate can be set to continue the infusion and maintain the venous
access after the completion of the delivery of the programmed volume to be infused.
The KVO rate can be preset in configuration menu. The actual KVO rate used will be the
KVO rate or the basal rate whatever is the smallest. The KVO will last for 20 ml volume.
Note: If this function is not required, it can be disabled. Refer to the Chapter 9, paragraph
9.2 "Configuration Menu ".
Automatic restart of the infusion after occlusion release.
The infusion will be automatically resumed following an occlusion release, this happens
when the pump has detected that the pressure is lower than the occlusion alarm level.
Note: If this function is not required, it can be disabled. Refer to the Chapter 9, paragraph
9.2 "Configuration Menu ".
Bolus rate
The rate to deliver bolus can be set in Configuration menu. The bolus rate can be set
in a value between 50-100 ml/h.
Warning before end of infusion
The warning can be set either for remaining Volume or for remaining Time. If Volume is
selected for warning, then it can be set from 1 to 10 ml inc. by 1 or it can be turned OFF.
If Time is selected for warning, it can be set from 5-180 min or it can be turned OFF.
Note: In Bolus Only mode the Time to warn before End Of Infusion cannot be used. In case
pump is configured with Time to warn before End Of Infusion and a Bolus Only mode is
selected then the pump will automatically enable the Volume to warn before End Of Infusion
at the default value (3 ml).
Note: If this function is not required, it can be disabled. Refer to Chapter 9, paragraph 9.2
"Configuration Menu ".
Protection against infusing without priming (Prime)
There are two (2) levels of protection that can be configured. First level is Mandatory;
infusion cannot start unless priming with the pump is done and a relevant message
will be displayed on the screen. Second level is Warning, infusion can start without
priming, a message appears on the screen, press Yes to continue with infusion.
Note: If this function is not required, choose No warning. Refer to Chapter 9, paragraph 9.2
"Configuration Menu ".