11. Precautions & Maintenance
11.1 Routine Maintenance Procedures
!Warning
Equipment can be modified as described in service manual by trained service personnel
with valid certificate. Appropriate inspection and testing according to service manual, must
be conducted to ensure safe use of the equipment.
There are no user serviceable parts or user replaceable parts within the pump, except
9V alkaline disposable battery.
Do not attempt to service the pump, only qualified service personnel can perform service
/ preventive maintenance.
!Warning
Ensure that the pump is OFF and disconnected from any power supply before
commencing the cleaning / disinfection procedure.
In order to avoid cross-contamination, for each equipment use a new clean, lint- free
cloth and pair of gloves and discard them after use, according to local protocols.
During cleaning / disinfection procedure, pay attention that disinfectant solution
does not go inside connectors.
Always inspect the battery terminals for debris, grime and damages. If debris or grime
exists, clean the terminals according to cleaning instructions. If the debris or grime per-
sists or damage exists remove the device from service. Debris, grime or damage of the
battery contacts could result in loss of power and non delivery of medication.
Contact your healthcare professional regarding any required service.
All servicing should be performed by a qualified service engineer with reference to the
MICREL Service Manual for this product. A comprehensive service manual containing
circuit descriptions, servicing and testing information is available for this unit. It can be
ordered from Micrel Medical Devices or Micrel Medical Devices authorized distributor.
Only trained technicians can perform service / preventive maintenance. In order to
send the equipment for service / preventive maintenance, place the pump with its
accessories in the original package, if available.
Preventive maintenance must be performed every 3 years or 1000 days of infusion and
only by authorized service personnel.
11.2 Cleaning and disinfection