86 87
!Warning
•
Ensure that the IP Connect Pack is OFF and disconnected from the mains supply and
from the pump before commencing the cleaning / disinfection procedure.
•
Ensure that the bolus handset and the power adaptor are disconnected from the
pump and from the mains supply (for power adaptor) before commencing the cleaning /
disinfection procedure.
11.2.3 Cleaning and disinfection procedure for accessories
1. Bolus handset, IP connect Pack and power adaptor
1. Wear a new pair of protective sulphur free gloves and eye protection if splashing is
likely to occur.
2. Slightly damp a clean, soft, lint -free cloth (microfiber) with solution of mild soap
(pH 7-8) and warm water (up to 30°C/86°F) and wring thoroughly. Make sure that it
is not dripping.
3. Do not clean the area under the battery cover for IP Connect Pack, inside of the
connector at the end of the cable for both IP Connect Pack and Micrel Power adaptor
and the metal parts of the plug.
4. Wipe the surface to dry using a clean, soft, lint-free cloth damped with water.
5. Let air dry. Do not use compressed air to dry the accessories.
6. Use one of the recommended disinfectant solutions listed in paragraph 11.2.2 to
spray on a clean, soft, lint-free cloth (microfiber). The cloth must be slightly damp
with the sprayed disinfectant.
7. Wipe the accessory using the damped cloth and repeat wiping for 5min.
8. Do not open the battery cover of the IP connect Pack to clean inside.
9. Be careful not to clean inside the connectors and the metal parts of the plug.
10. Wipe the surface using a clean, soft, lint-free cloth damped with water.
11. Let air dry. Do not use compressed air to dry the accessories.
After removing protective equipment on completion of task, thoroughly wash and dry
hands. The disposable equipment must be discarded according to local regulations.
2. Carrying bag (except disposable carrying bags, they are single use)
In order to clean the carrying bag, please follow the below procedure:
1. Wash carrying bag with cold water either in wash machine or hand wash.
2. Do not use bleach.
3. Dry it in shade, do not use dryer and do not wring.
4. Do not iron the carrying bag.
In order to disinfect the carrying bag, please follow the below procedure:
1. Wear a new pair of protective sulphur free gloves and eye protection if splashing is
likely to occur.