5.2.6 Programming procedure Bolus only mode
42 43
Enter the patient
BOLUS
dose value and press .
Select one of the two limits by pressing the appropriate
key.
Enter the
LOCKOUT INTERVAL
value in minutes and press
If this parameter is not required, the value 000 should be entered.
BOLUS
05.0 m1
LOCKOUT INTERNAL
015 min
SELECT LIMIT
DOSE BOLUS
LIMIT / HOURS
Enter the DOSE LIMIT value and press .
DOSE LIMIT
observation window can be changed in this screen
by pressing the
TIME
soft key.
MAX BOLUS in the dose limit screen displays the maximum
possible number of patient bolus that the patient can get
within the preset period of time.
or
Enter the patient BOLUS LIMIT value and press .
BOLUS LIMIT
observation window can be changed in this screen
by pressing the
TIME
soft key.
Note: If dose limit or bolus limit is not required, the value 0000
should be entered.
2 MAX BOLUS
DOSE
LIMIT
0180 m1/4hr
TIME
BOLUS
LIMIT
010 BOL /4hr
TIME
VOL TB INF 160.0 m1
BOLUS 5.0 m1
LOCKOUT 15 min
LOAD DOSE 2.0 m1
BOLUS / H 10BOL /4hr
At the end of programming a summary of the protocol appears
on the screen, the programmed value of each parameter is
displayed.
Press to scroll down the protocol.
!Warning
Always validate infusion protocol values by checking all parameters, prior to start the
infusion. If the Protocol values are not correct, please, restart the programming
procedure by pressing when you are on the "PUMP ON HOLD" screen.
Enter the LOADING DOSE value if required and press
Note: If this parameter is not required, the value 000.0 should
be entered.
LOADING DOSE
02.0 m1
Enter the VOLUME TO BE INFUSED value and press .
MAX/24H volume indication at the top of the screen indicate
the maximum volume that a patient can receive within 24
hours, according to the above programmed values of each
parameter.
MAX/ 24H 1010m1
VOLUME TO BE INFUSED
0160 m1
.
.