!Warning
•
Periodically inspect the fluid path for air. If air is seen stop infusion. Air infusion could
result in under or non-delivery of medication. Air embolism could result in death or serious
injury to the patient.
•
Periodically inspect the fluid container for decreasing volume, inspect the fluid path for
kinks, a closed clamp or other upstream obstruction. Upstream occlusions could result in
under or non-delivery of medication.
Infusion has started. The programmed parameters and their
perspective values of the running protocol can be reviewed
at once using the
PROTOCOL
key.
46 47
5.2.8 Pre-programming the pump
It is possible to program the pump in advance and turn it Off after programming. The
protocol is then stored in the pump. When the pump is ready and the set fitted, Turn
the pump on and:
CODE:
0000
Enter resuming” code and press .
If Protocol is correct, press twice to start the
infusion.
11 APR 2014 11:45
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
PGM
11 APR 2014 11:45
PUMP RUNNING
HISTORY
PROTOCOL
!Warning
Always validate infusion protocol values by checking all parameters, prior to start the
infusion. If the Protocol values are not correct, please, restart the programming
procedure by pressing when you are on the "PUMP ON HOLD" screen.