6.9 Answering to Pain level question
Pain level question can be manually accessible as described below:
11 APR 2014 11:45
PUMP RUNNING
HISTORY
PROTOCOL
Press when the pump is on hold or during infusion.
Press
YES
to enter the questionnaire.
Press the – or + soft keys to select value in the 1-10 scale.
Press to answer.
After finish answering to question, the following message will
be displayed, informing that the answer is recorded.
DO YOU WANT TO
ANSWER QUESTIONS ?
NO YES
THANK YOU ,
YOUR ANSWERS
WERE RECOVERD
PAIN LEVEL
2
6.10 Viewing the remaining volume to be infused
and the infusion parameters.
When the pump is on hold or during infusion,
11 APR 2014 11:45
PUMP RUNNING
HISTORY
PROTOCOL
Press the
PROTOCOL
key to review the current infusion
parameters.
Press to view the second screen.
Note: when KVO is running the KVO rate is displayed instead
of the basal rate value.
Note: when programmed or/and if the pump is connected
to the web the patient name, birth date and protocol name
are displayed.
Press to return to the main screen.
Volume TB infused is the remaining volume to be infused. Press to return to
the main screen.
PUMP ASSIGNED TO:
Mr PHIL. HIPPAIN
BIRTHDATE: 27 / 06 / 58
PROTOCOL: MORPHINE I.V. 10mg
VOL TB INF 160.0 ml
BASAL RATE 5.0 ml / hr
VOL TB INF 160.0 ml
BOLUS 5.0 ml
LOCKOUT 15 min
LOAD DOSE 2.0 ml
BOLUS/H 10BOL /4HR
BOLUS 5.0 ml
LOCKOUT 20 min
DOSE LIMIT 195ml /4hr
VOL TB INF 160.0 ml
BASAL RATE 40.0 ml / hr
AUTO - BOLUS 5.0 ml
FREQUENCY 0h30’
LOAD DOSE 2.0 ml
e.g. Continuous mode
e.g. Bolus only mode
e.g. Auto Bolus mode