12.6 User Interface Alarms & Warnings
• Occlusion detection
• End of infusion / Programmed volume limit reached
• Depleted battery
• Open mechanism
• Program error
• Technical error
• Dose limit alarm
• Bolus limit alarm
• Near end Of infusion / Pre-end of programmed volume or time limit
• Low battery
• IP connect Pack low battery
• No patient bolus allowed
• Running KVO
• Dose limit reached. Bolus not allowed
• Preventive maintenance date
• Real time clock battery low
• Keyboard continuously pressed
• Value out of range
• Stand by audible buzzer sound every minute
94 95
12.5 Performances
Memory retention 10 years
Expected service life 10 years
Maximum over infusion volume under single fault condition 0.5 ml
Max Accuracy error (pump speed) under single fault condition
± 5%
Rate accuracy in all available range remains within acceptable
limits when:
temperature range
+ 5°C to +40°C
delivery rate range
0.1 ml/h – 100 ml/h
0.1 to 999.9 mg/h
0.1 to 999.9 μg/h
Max time to occlusion PL1 @ 1 ml/h
@ 25.0 ml/h
@ 100.0 ml/h
Max time to occlusion at PL3 @ 1 ml/h
@ 25.0 ml/h
280 sec
10 sec
4 sec
380 sec
17 sec
Bolus at occlusion release at PL1 @ 25.0 ml/h
Bolus at occlusion release at PL1 @ 25.0 ml/h
0.1 ml max