12.6 User Interface Alarms & Warnings
Occlusion detection
End of infusion / Programmed volume limit reached
Depleted battery
Open mechanism
Program error
Technical error
Dose limit alarm
Bolus limit alarm
Near end Of infusion / Pre-end of programmed volume or time limit
Low battery
IP connect Pack low battery
No patient bolus allowed
Running KVO
Dose limit reached. Bolus not allowed
Preventive maintenance date
Real time clock battery low
Keyboard continuously pressed
Value out of range
Stand by audible buzzer sound every minute
94 95
12.5 Performances
Memory retention 10 years
Expected service life 10 years
Maximum over infusion volume under single fault condition 0.5 ml
Max Accuracy error (pump speed) under single fault condition
± 5%
Rate accuracy in all available range remains within acceptable
limits when:
temperature range
+ 5°C to +40°C
delivery rate range
0.1 ml/h – 100 ml/h
0.1 to 999.9 mg/h
0.1 to 999.9 μg/h
Max time to occlusion PL1 @ 1 ml/h
@ 25.0 ml/h
@ 100.0 ml/h
Max time to occlusion at PL3 @ 1 ml/h
@ 25.0 ml/h
@ 100.0 ml/h
280 sec
10 sec
4 sec
380 sec
17 sec
4 sec
Bolus at occlusion release at PL1 @ 25.0 ml/h
Bolus at occlusion release at PL1 @ 25.0 ml/h
0.1 ml max