76
In case of use the Micrel power adaptor with Rythmic™ Evolution with 9V Alkaline
battery, Micrel power adaptor will not recharge the battery.
The alkaline battery must be in good condition as it assures continuation of infusion,
in case of Mains power failure.
The use of the Micrel power adaptor does not inhibit the power of the alkaline battery
and will not cancel any battery related warning or alarms. Micrel power adaptor is
recommended to extend the life of internal battery and avoid battery alarms at night.
9V Alkaline battery may last for a year if Micrel power adaptor is always in use.
The unit does not require any maintenance and must not be opened or manipulated/
changed in any other way.
10.3 Compatible administration set
Below is a list with the approved Administration sets and also their respective pump to be used
with. The delivery accuracy of the below Rythmic™Administration sets is less than ±5%.
Code Description
Rythmic™
Evolution
Mini
Rythmic™
Evolution
Rythmic™
Evolution
Organiser
Not for
IV use
Full set
160ml
KM1.EE.100.x
Complete set with 160ml bag with filling port, air
in line eliminating filter 0.22micron, antisiphon /
back check valve and extension line.
x
Full set
160ml 1.2u
KM1.EE.135.x
Complete set with 160ml bag with filling port,
air in line eliminating filter 1.22micron, antisi-
phon /back check valve and extension line.
x
!Caution
Always keep the internal battery (9V alkaline or rechargeable) inside the pump while the
Micrel power adaptor is plugged to the pump.
All Rythmic™ pumps can be powered using Micrel power
adaptor in addition to the internal battery (9V alkaline or
rechargeable). Please refer to Chapter 10 paragraph 10.7
“Accessories”. Micrel power adaptor is part of the Rythmic™
Evolution system.
Place the power supply to a mains socket and plug the cable
end into the socket labelled on the pump.
When the power adaptor is connected to the mains supply, the green LED indicator
lights up.
When operating on mains, the mains symbol is displayed nearby the battery
indication .
During operation, the housing of power adaptor may heat up.