Scroll down the screens by pressing until you reach the para-meter you need to change.
Modify this parameter and press .
Scroll down to the “PUMP ON HOLD” screen and Press
twice to resume the infusion.
11 APR 2014 11:45
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
Titr
The clinician bolus override function provides the medical staff with the capability to
administer additional bolus doses to the patient, even during the lockout interval.
The clinician bolus dose is included in the dose limit counter totals.
6.7 Clinician Bolus override
Press
PROGRAM
.
56 57
!Warning
Extra care when using the clinician bolus function. There is no limit, or frequency limit in
the use of the override bolus function. Incorrect programming could result in death or serious
injury to the patient.
Do not leave the pump unattended while is on Clinician Bolus screen.
We recommend the default codes of the pump to be changed (through the configu-
ration menu) in order to secure the access to different actions related to each code.
The supervisor of the infusion should decide to whom each code and its related actions
should be distributed. Incorrect pump programming could result in death or serious injury to
the patient.
CODE:
0000
11 APR 2014 11:45
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
STOP
Put the pump on hold by pressing twice and enter the
code to stop infusion, if code protected stop of infusion is
enabled. Press .
11 APR 2014 11:45
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
STOP
Enter your “programming” code and press .
Select
CLINIC. BOLUS
.
11 APR 2014 11:45
PUMP ON HOLD
PROGRAM
Titr
CLINIC. BOLUS
The Clinician bolus dose is pre-set to the value of the normal
patient bolus dose. If needed adjust the dose and press
.
CLINICIAL BOLUS
02.0 ml
Pressing twice will infuse the Clinician bolus dose and
resume the infusion.
11 APR 2014 11:45
PUMP RUNNING
HISTORY
PROTOCOL

Table of Contents