Volume to be infused 1 to 9999 ml 1 ml increment (inc)
Volume infused
0.1-9999.9 ml 0.1 ml inc.
0.1 mg to 9999.9 mg 0.1 mg inc.
0.1 μg to 9999.9 μg 0.1 inc.
Programming units
ml/h; mg/h; μg/h
Concentration
0.1-99.9 mg/ml 0.1 mg/ml inc.
0.1-999.9 μg/ml 0.1 μg/ml inc.
Basal Rate
From 0 to 100 ml/h 0.1 ml/h inc. (0 to 50 ml/hr
when bolus is on), 0.1 to 999.9 mg/h 0.1 mg/h inc.
0.1 to 999.9 μg/h 0.1 μg/h inc.
Rate Accuracy ±5% nominal
KVO 0.1 to 2.0 ml/h KVO limited in volume to 20 ml.
Automatic Bolus Dose
0 or from 0.1 ml to 60.0 ml 0.1 ml inc.
0.1- 999.9 mg 0.1 mg inc.
0.1- 999.9 μg 0.1 μg inc.
Automatic Bolus Frequency 0 to 9 h 59 min 1min inc.
Bolus Dose
0 or from 0.1 ml to 60.0 ml 0.1 ml inc.
0.1- 999.9 mg 0.1 mg inc.
0.1- 999.9 μg 0.1 μg inc.
Bolus accuracy ±5% nominal according to IEC 60601-2-24
Bolus Lockout interval
0 to 9 h 59 min 1 min inc.
Priming Rate
Automatic adjustable, maximum 100 ml/hr
Bolus Rate From 50 to 100 ml/h, 10 ml/h inc.
Dose Limit Volume
OFF or from 0 to 9999 ml 1ml inc.
or 9999 mg 1 mg inc. or 9999 μg from 0.1-99.9 μg/ml 1μg
inc. and 99990 μg from 100.0 - 999.9 μg/ml 10 μg inc.
Bolus limit Off or from 0 to 999 boli 1 bol inc. limited to 100 ml/h
Dose Limit & bolus limit
interval
1-9 HR 1 hr inc.
12.3 Performance & Protocol parameters