11.2.1 Cleaning procedure for pump
To insure that the equipment remains in good operating condition, it is important to keep
it clean. Clean the pump periodically following the below described procedure.
1. Wear a new pair of protective sulphur free gloves and eye protection if splashing
is likely to occur.
2. Slightly damp a clean, soft, lint -free cloth (microfiber) with solution of mild soap
(pH 7-8) and warm water (up to 30°C /86°F) and wring thoroughly. Make sure that it
is not dripping.
3. Wipe the pump using the damped cloth. Do not clean the area under the tubing
cover. This area must be cleaned by special trained personnel.
4. Slide the battery door and remove the battery. Wipe the battery compartment.
5. Scrub and clean the surface of the battery terminals using a clean, soft, lint-free cloth
(microfiber) with lightly damped with isopropyl alcohol. Do not use isopropyl alcohol
to clean other parts of the device than the battery terminals
6. Wipe the surface to dry using a clean, soft, lint-free cloth damped with water.
7. Let air dry. Do not use compressed air to dry the pump.
11.2.2 Disinfection procedure for pump
In order to disinfect the pump, after cleaning procedure please follow the below described
procedure. Disinfection must take place regularly and always when the user is changed.
8. Use one of the recommended disinfectant solutions listed below to spray 2 or 3
times on a clean, soft, lint-free cloth (microfiber). The cloth must be slightly damp
with the sprayed disinfectant.
9. Wipe the pump using the damped cloth and repeat wiping for 5min.
10. Open the transparent covers and wipe the pump inside. Do not open the tubing
cover to disinfect inside. This area must be cleaned by special trained personnel.
11. Wipe the surface to dry using a clean, soft, lint-free cloth damped with water.
12. Let air dry. Do not use compressed air to dry the pump.
After removing protective equipment on completion of task, thoroughly wash and dry
hands. The disposable equipment must be discarded according to local regulations.
Wipe Information
material and weight: Non-woven PET (30%) / viscose (70%) blend, 50g/m2
size: 100 x 150 mm or
100 x 200 mm