62 63
7.5 Infusion events log overview
Infusion events are only accessible through uploading event log to PC. These include:
User Interface warnings, User Interface Alarms, Pump On/Off events, Infusion Start/
Stop, Priming Start/Stop, Loading Dose, Bolus Start/Given/Demanded, Titration, New
Bag and Resume after power off.
7.6 Viewing infusion data
even after turning off
the pump
If the pump has been turned OFF, it is possible to review the above mentioned infusion
data/graphs by turning ON the pump again and entering the “resume” code.
If however a “new patient-program” function has been selected after turning ON, the
infusion data of the previous infusion will only be available through uploading functions.
11 APR 2014 11:45
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
STOP
7.7 Uploading event log to a computer
All the data can be uploaded to a PC for archiving purposes or custom statistical
analysis. To upload event log to a computer, the “Rythmic™ Data Manager” software
is required. Rythmic™ Data Manager is part of the Rythmic™ Manager Pack reference
KP5.04.250.x. Data can be saved in a computer either in an excel format or in text
format.
!Warning
•
The pump should be disconnected from the patient prior to being connected to a
computer.
•
Do not perform this operation in the presence of a patient.