8. User Interface Alarms and Warnings
The Rythmic™ Evolution pump has means to make the user aware of any eventual
equipment malfunction, and/or wrong operator equipment usage, and/or abnormal
physiological conditions. There are three categories: User Interface Warnings, User
Interface Alarms and Technical Alarms.
8.1 User Interface Warnings
Whenever user interface warnings are produced, the infusion continues and an
identification message is displayed. Most of the user interface warnings have audible
and visual signals. In case of audible warnings the pump sounds intermittently, press
the key to silence the acoustic tone for one hour, but maintain the warning
message on the display.
The sound alarm of the near end of infusion warning can be disabled on the pump to
be forwarded by SMS messages when IP connect Pack is used. Refer to Chapter 10,
paragraph 10.4 "IP Connect Pack". Press to acknowledge.
The first time this warning is triggered, the remaining battery life is at least 12 hours.
Near end of Infusion
Low Battery
IP connect Low Battery
Audible YES
Visual YES
Acoustic tone:
Intermittent
Audible YES
Visual YES
Acoustic tone:
Intermittent
Audible YES
Visual YES
Acoustic tone:
Intermittent
Cause
The remaining volume
to be infused is low.
Infusion will end soon.
Cause
Battery power is low
but the infusion is still
on going.
Cause
The battery of
accessory IP Connect
Pack is low. Infusion
will not stop but
connection might be
lost.
Action
Press to acknow-
ledge the message.
Anticipate on next action
to be taken.
Action
Change the battery when
the infusion is finished.
Action
Charge the battery of
IP Connect Pack, as
instructed in Chapter
10 paragraph 10.4 “IP
Connect Pack.”