No patient bolus allowed
Audible NO
Visual YES
Cause
A patient bolus is
demanded but not
allowed eg: during
bolus infusion,
during lockout time
or because dose or
patient bolus limits
have been reached.
Action
Wait for the time limit
to pass. Patient bolus
will automatically be
authorized after the limit
elapsed.
Type of signal
Type of signal
Type of signal
Type of signal
Type of signal
Warning message
BOLUS NOT
AUTHORISED
BEFORE 0'20''
Running KVO
Audible YES
Visual YES
Acoustic tone:
Continuous
Cause
The remaining volume
to be infused has
been delivered and
pump started infusing
at KVO rate.
Action
If necessary, leave the
pump to run with KVO
rate. Or, turn the pump
OFF. When the maximum
KVO volume will be
reached the infusion will
stop and an alarm will be
triggered.
Warning message
RUNNING KVO
VOLUME TO BE
INFUSED
DELIVERED
DOSE LIMIT
REACHED BOLUS
NOT ALLOWED
Dose limit reached. Bolus not allowed
Audible NO
Visual YES
Cause
The requested patient
bolus exceeds the
dose limit.
Action
Wait for the limit to be
elapsed, patient bolus
will automatically be
authorized.
Warning message
Preventive maintenance date
Audible NO
Visual YES
Cause
There have been
passed 3 years
or 1000 days of
infusion. The date for
preventive inspection
is approaching.
Action
Press to
acknowledge the
message. Return the
pump to a qualified
technician for routine
preventive inspection
when the pump is
available.
Warning message
ROUTINE
MAINTENANCE TO
BE PERFORMED
COD
Low real time clock battery
Audible NO
Visual YES
Cause
A problem detected
with the internal real
time clock battery.
Action
Return the pump to a
qualified technician to
change the battery.
Warning message
REAL TIME
CLOCK BATTERY
LOW