Put the pump on hold by pressing twice and then
press
PROGRAM.
.
Enter new bag code and press .
Repeat the procedure described above in order to enter the new Volume to be Infused and to
start infusion.
6.4 Resuming the same infusion with the same patient
after power off of the pump
(Resume after power off)
This function allows continuing the infusion at the stage it was before power OFF,
e.g. when 9V alkaline battery is replaced. There is no ability to change any infusion
parameters. Since the infusion has already started before power OFF, there is no
ability to prime the administration set using the pump.
The user can access this feature after turning the pump ON and either enter Resume
after power off code or enter the New patient – program code and select “Resume
after power off” from selection menu.
54 55
During infusion, user can access “Change bag” following the below procedure:
CODE:
0000
!Warning
•
Ensure that air has been purged from the administration set before connecting it to
patient, to prevent air embolism. Air embolism could result in death or serious injury to the
patient.
•
Ensure that the administration set is properly installed and has not been trapped in
transparent covers. An undetected upstream occlusion may result in under or non-delivery
of medication. Depending upon medication under or non-delivery could result in death or
serious injury to the patient.
•
Ensure that the release lever is locked in closed position and the tubing cover is closed
firmly. If the release lever or the tubing cover become loose, may result in under on non-
delivery of medication.
•
Periodically inspect the fluid container for decreasing volume, inspect the fluid path for
kinks, a closed clamp or other upstream obstruction. Upstream occlusions could result
in under or non-delivery of medication, which could result in death or serious injury to the
patient.
Use to select “Resume after power off”.
Press .
The below sequence is the same when using programming code and new bag code.
NEW PATIENT -PROGRAM
NEW BAG
RESUME AFTER POWER OFF
11 APR 2014 11:45
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
STOP