!Warning
Always validate infusion protocol values by checking all parameters, prior to start the
infusion. If the Volume to be infused value is not correct, please, modify it by pressing
to get on the relevant screen. Wrong infusion protocol could result in death or serious
injury to the patient.
11 APR 2014 11: 22
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
Titr
If the Volume to be infused is correct, press
twice to start the infusion.
!Warning
Ensure that air has been purged from the administration set before connecting it to patient,
to prevent air embolism. Air embolism could result in death or serious injury to the patient.
Do not prime the administration set while it is connected to a patient, as this may result in
air embolism or over- delivery of medication. Always disconnect the line from patient before
priming.
HAVE YOU
PRIMED THE SET?
STOP
NO
YES
Make sure that the administration set is fully primed. If not,
press
NO
and prime the set using the pump, refer to Chapter
6, paragraph 6.1 "Priming the Administration set line". If the
administration set is primed, press
YES
.
!Warning
Periodically inspect the fluid path for air. If air is seen, stop infusion. Air infusion could
result in under or non-delivery of medication. Air embolism could result in death or serious
injury to the patient.
Periodically inspect the fluid container for decreasing volume, inspect the fluid path for
kinks, a closed clamp or other upstream obstruction. Upstream occlusions could result in
under or non-delivery of medication.
Protocol’s parameters are shown in summary.
Press to scroll down the protocol.
54 55
11 APR 2014 11:45
PUMP RUNNING
IV MORPHINE
HISTORY
PROTOCOL
Infusion has started. The programmed parameters and their
perspective values of the running protocol can be reviewed
at once using the
PROTOCOL
key
.
VOL TB INF 160.0 m1
BOLUS 5.0 m1
LOCKOUT 15 min
LOAD DOSE 2.0 m1
BOLUS / H 10BOL /4hr