!Warning
Always validate infusion protocol values by checking all parameters, prior to start the
infusion. If the Protocol values are not correct, please, restart the programming
procedure by pressing when you are on the "PUMP ON HOLD" screen.
40 41
VOL TB INF 160.0 m1
BASAL RATE 42.1 m1/hr
If Protocol is correct, press twice to start the
infusion.
At the end of programming a summary of the protocol
appears on the screen, the programmed value of each
parameter is displayed. Press to scroll down the
protocol.
11 APR 2014 11:45
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
PGM
Make sure that the administration set is fully primed. If not,
press
NO
and prime the set using the pump, refer to Chapter
6, paragraph 6.1 "Priming the Administration set line". If the
administration set is primed, press
YES
.
HAVE YOU
PRIMED THE SET?
STOP
NO
YES
!Warning
Ensure that air has been purged from the administration set before connecting it to
patient, to prevent air embolism. Air embolism could result in death or serious injury to the
patient.
Do not prime the administration set while it is connected to a patient, as this may result in
air embolism or over- delivery of medication. Always disconnect the line from patient before
priming.
!Warning
Periodically inspect the fluid path for air. If air is seen stop infusion. Air infusion could
result in under or non-delivery of medication. Air embolism could result in death or serious
injury to the patient.
Periodically inspect the fluid container for decreasing volume, inspect the fluid path for
kinks, a closed clamp or other upstream obstruction. Upstream occlusions could result in
under or non-delivery of medication.
Infusion has started. The programmed parameters and their
perspective values of the running protocol can be reviewed
at once using the
PROTOCOL
key.
11 APR 2014 11:45
PUMP RUNNING
HISTORY
PROTOCOL