Volume to be infused
The volume to be infused parameter is used to trigger the near end of infusion alarm
and then later to trigger the end of infusion alarm.
Usually the volume to be infused value is the bag volume minus the primed volume.
The volume to be infused value, displayed on the pump screen, will decrease during
the course of the infusion whilst the infused volume will increase accordingly.
Protocol library
The user can transfer customised protocol library from the PC application “Rythmic™
Evolution Therapy Manager” to the Rythmic™ Evolution pump. The protocol
library contains preconfigured protocols, for which the values of the programmed
parameters are pre-set, with or without limits in order to achieve programming errors
prevention.
For detailed information regarding the operations and configuration for uploading a
protocol library to the pump please refer to “Rythmic™ Manager Pack DFU”.