B
A
5. Load the tubing into pumping mechanism. It can only be loaded one way
round. Check that the pressure sensor is located under the door and
positioned with the membrane facing downwards. The locating key should be
correctly orientated. The tube shall exit the pump and connect to the patient
side. Please make sure the tubing and the bag are placed properly inside the
transparent covers.
A
B
C
D
7. Close the transparent covers. Ensure the transparent covers have been locked
in position.
8. Unclamp the set, before priming through the pump or starting the infusion.
C
D
6. Close the tubing cover firmly.
!Warning
Ensure that the transparent covers are securely closed, to prevent patient access to
medication container.
Ensure that the administration set has been properly installed and has not been trapped
in transparent covers, as this may result in under or non-delivery of medication.
Ensure that the release lever is locked in closed position and the tubing cover is closed
firmly. If the release lever or the tubing cover become loose may result in under or non-
delivery of medication
.