For sets with Spike and external medication container, Rythmic™ Spike Sets
Rythmic™ Spike Sets can fit in all models of Rythmic™ Evolution range.
1. Open the transparent cover.
2. Press the tubing cover release lever to open the cover. This will expose the
pumping mechanism.
3. Remove empty administration set, if appropriate.
4.
Load the tubing into pumping mechanism. It can only be loaded one way
round. Check that the pressure sensor is located under the door and
positioned with the membrane facing downwards. The locating key should
be correctly orientated, inserted in the appropriate space in the battery door and
pushed firmly until it clicks . The tube shall exit the pump and connect to
the patient side.
B
C
A
D
26 27
!Warning
Ensure that the slide clamp is closed before removing the administration set from the
pump, to prevent free flow of medication.
Always inspect for marks, dust, dirt, particles or substance that exist underneath the
tubing cover or in the mechanical pumping area before use. Any of the above could result
to under or non-delivery of medication. The pump should be immediately taken out of use
and inspected by qualified service personnel to ensure its proper function prior to reuse.
A
For Rythmic™ Organiser:
Lift the door hood to open. Press the tubing cover
release lever to open the cover. This will expose
the pumping mechanism.