Before use the pump and it’s accessories to carry out an infusion, run the following
test for a complete alarm and safety features check for both pump and accessories.
Pump’s Serial number (S/N): .............................................. Date: ..... / ..... / .....
Accessory information (description) : …………………........................................…….
If the device fails to perform as described in one or more of the above tests, pump
and its accessories should be immediately taken out of use and inspected by
authorized service personnel to ensure their proper function prior to use.
18 19
Visual inspection for Pass Fail
Absence of impact marks / broken covers / damaged wires
Sufficient legibility of all labels
Absence of dust, dirt, particles or substances underneath the tubing cover
Absence of liquid ingression / battery compartment
Absence of debris, grime or damage of battery terminals
Functional test
Locking system operating properly
Action Result Pass Fail
Test result :
While power adaptor is connected
to the Rythmic™ Evolution remove
disposable alkaline battery and
start infusion.
“BATTERY NEEDED AS BACK UP”
message should appear and
infusion should not start.
Turn the pump ON and verify
that buzzer functions properly.
An acoustic tone will sound
twice.
Check display during power ON.
No missing dots on dark
display. Backlight turns ON.
Check for correct date / time. Actual date / time.
Connect power adaptor to Mains
and to Rythmic™ Evolution.
Mains indicator should
appear.
Test performed by :
Department: ………………………. Name: …………………………………………….
“DOOR IS OPEN” alarm should
appear. Press to silent the
acoustic tone.
“DOOR IS OPEN” alarm should
appear. Press to silent the
acoustic tone.
Enter the new bag code, enter 1ml
as volume to be infused and then
press to reach “PUMP ON
HOLD” screen. Press to
prime (without administration set
inside).
Press to start infusion
(without administration set inside).