!Cautions
Do not clean the equipment using acetone, other plastic solvents or abrasive clean-
ers as they may damage the equipment.
Do not use hard or pointed objects to clean any part of the pump and its acces-
sories.
Do not spray cleaning fluids directly on the pump and its accessories.
Do not steam autoclave, ethylene oxide sterilise or immerse Rythmic
Evolution pump and its accessories in any fluid.
Do not use UV radiation to disinfect the pump and its accessories.
Do not use compressed air to dry the pump and its accessories.
Do not mix the disinfectant with any other product or chemical.
Do not immerse the pump in cleaning fluid or water, it is not waterproof and it can
cause serious damage to the pump.
Do not store the pump with the battery inside for more than three months.
Do not store the pump with the rechargeable battery depleted. If the pump has to
be stored for a long period of time, every three months recharge the battery.
Check that the polarity of the 9V alkaline battery is correctly oriented.
Check that the battery door has been closed correctly, otherwise the transparent
covers may not close.
Do not remove the battery while the pump is switched ON as this may damage the
pump.
Prior to using IP Connect Pack accessory for the first time, use the power adaptor
to fully charge the internal rechargeable battery.
Do not remove the internal battery (9V alkaline or rechargeable) from the
pump while it is connected to Micrel power adaptor or IP Connect Pack accessory.
Prior to using pump with rechargeable battery for the first time, use the power adap-
tor to fully charge the internal rechargeable battery.
• During charging of Rythmic™ Evolution while the pump is switched OFF, there is not
any indication of charging progress.
When handling fluid-path connections always follow aseptic technique protocol.
Tighten the luer connectors to prevent leakage.
The administration set is a single use product and must be changed in accordance
with the applicable healthcare protocols.
Prior to administering medication under specific delivery conditions, medication
stability for time and temperature must be checked.
The use of non reccomended cleaning solutions and disinfectants and the failure to
follow Micrel cleaning and disinfection procedure may result in product damage.