Chapter 12 Serial Bus Trigger and Decode (Optional)
127
Trigger mode selection menu description:
a) F.start — trigger at the start of the frame;
b) Remote ID setting the ID matches the remote frame trigger. After selecting the “Remote Frame ID”,
and then set the ID value at the bottom of the trigger data area
Operation description: Press the numbers on the touch screen and use the virtual keyboard to set;
c) Data ID trigger on data frame that matches set ID. Data frame ID configuration mode is the same
as the remote data frame ID configuration;
d) R/D ID trigger on remote frame or data frame that matches set ID. Remote frame/data frame ID
configuration is the same as the remote data frame ID configuration;
e) ID + data — trigger on data frame that matches set ID and data. The configuration method is the same
as the remote frame ID configuration;
f) Wrong F. — trigger on CAN error frame;
g) All Error — trigger when there is any error in format or activity;
h) Ack Error — trigger on recessive (high) Ack position;
i) Over Load — trigger on CAN overload frame.
CAN serial decode
Ch1 is connected to measured signal. The idle level is high and the baud rate is 1Mb/s; the Trigger mode is
the frame start. Please follow these steps:
(1) Tap S1 to open the decode channel, and click S1 again to open the bus configuration menu;
(2) Select the bus type as “CAN”, and then click “Ch1”, “Idle High” and “1Mb/s”. After setting, click the
blank area to close the menu;
(3) Open the trigger mode configuration menu and click “F.Start”;
(4) Adjust the threshold level according to the signal amplitude; the CAN trigger graphic interface is shown
in Figure 12-16: