Milli-Q® Direct 8/16 System 113
Summary of Alert messages, Continued
Alert messages
(continued)
(continued)
TAP FEED CONDUCTIVITY > SP
The Tap Water conductivity is below
the set point.
Contact Millipore SAS.
THE NETWORK CABLE IS
UNPLUGGED
Check the Ethernet Cable plugged
into the System and the computer.
Restart the System.
THIS IP ADDRESS IS ALREADY
USED BY ANOTHER SYSTEM
Contact your network administrator.
Restart the System.
TOC < 1 PPB
Invalid TOC measurement.
Perform a TOC Curve Check.
Contact Millipore SAS.
TOC FEEDWATER < 15.3 M
.cm
Due to low resistivity at the inlet of
the UV Lamp, the TOC measurement
can not be performed properly
anymore.
Please replace the Q-Pak® Pack to
ensure valid TOC indications.