Milli-Q® Direct 8/16 System 25
Tubing, cables and power cord, Continued
Feed water
tubing to pipe
Install the Inlet Strainer as shown here.
Connect one end of the feed water tubing to the Inlet Strainer.
Reservoir
connections
The tubings from the Water System Ports 3 and 4 are connected to the Reservoir as
shown here.
NOTE:
The valve where the tubing from Port 3 is connected must be opened.
Powering the
system
Open the feed water source.
Plug the power cord into the Water System.
Plug the power cord into a source of electrical power.
The Main LCD shows a series of start-up screens.
Continued on next page

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Milli-Q Direct 8 and is the answer not in the manual?

Milli-Q Direct 8 Specifications

General IconGeneral
Resistivity18.2 MΩ·cm at 25 °C
Total Organic Carbon (TOC)< 5 ppb
Pyrogens (Endotoxins)< 0.001 EU/mL
RNases< 1 pg/mL
DNases< 5 pg/mL
DisplayTouchscreen
Weight20 kg
Power Supply100-240 V, 50/60 Hz
Type of Water ProducedPure and ultrapure water
Flow RateUp to 2 L/min
Particulates<1 particle/mL (>0.2 µm)
Particles > 0.22 μm< 1 particulate/mL
Particle Size0.22 µm
ConnectivityUSB
Output Water TypeUltrapure
Dimensions504 mm x 332 mm x 467 mm (19.8 in x 13.1 in x 18.4 in)