Operator’s Manual of Anesthesia System 6 - 3
Operations Patient Setup
6.3 Patient Setup
6.3.1 Standby Mode
Some system functions do not work when the system is in the standby mode.
Figure 6-1 Standby Mode
NOTE: Whether to display the [Standby] or [Start Case] key on the soft key
field of the Main Screen shall be set by authorized service personnel.
Please contact Mindray Technical Support if necessary.
6.3.1.1 Enter Standby mode
When the [Standby] key is displayed on Soft key field of the Main Screen
1. Select the [Standby] key on the Main Screen.
2. Follow the on-screen prompts to enter Standby mode.
When the [Standby] key is note displayed on Soft key field of the Main Screen:
1. Set the Auto/Manual switch to the Manual position.
2. When using the EFCS, skip to Step 3.
When using the BFCS, turn the flow control knob of the BFCS clockwise to the mechanical stop
position.
3. Select the [Standby] key in the Manual mode.
4. Follow the on-screen prompts to enter Standby mode.
NOTE: After selecting the [Standby] key, you can set whether to [Restore
current prrofile settings] in the pop-up dialog box.
NOTE: When the system is in Standby mode and the Auto/Manual Switch is in
Manual position, the [Bypass] button in the Manual tab is disabled.
However, the [Alarms] button remains enabled and can be toggled to
On or Off.
WARNING: Entering Standby mode will stop ventilation and parameter monitoring.
Do not select Standby mode if the patient requires continuous ventilation.