Operator’s Manual of Anesthesia System 6 - 3
Operations Patient Setup
6.3 Patient Setup
6.3.1 Standby Mode
Some system functions do not work when the system is in the standby mode.
Figure 6-1 Standby Mode
NOTE: Whether to display the [Standby] or [Start Case] key on the soft key
field of the Main Screen shall be set by authorized service personnel.
Please contact Mindray Technical Support if necessary.
6.3.1.1 Enter Standby mode
When the [Standby] key is displayed on Soft key field of the Main Screen
1. Select the [Standby] key on the Main Screen.
2. Follow the on-screen prompts to enter Standby mode.
When the [Standby] key is note displayed on Soft key field of the Main Screen:
1. Set the Auto/Manual switch to the Manual position.
2. When using the EFCS, skip to Step 3.
When using the BFCS, turn the flow control knob of the BFCS clockwise to the mechanical stop
position.
3. Select the [Standby] key in the Manual mode.
4. Follow the on-screen prompts to enter Standby mode.
NOTE: After selecting the [Standby] key, you can set whether to [Restore
current prrofile settings] in the pop-up dialog box.
NOTE: When the system is in Standby mode and the Auto/Manual Switch is in
Manual position, the [Bypass] button in the Manual tab is disabled.
However, the [Alarms] button remains enabled and can be toggled to
On or Off.
WARNING: Entering Standby mode will stop ventilation and parameter monitoring.
Do not select Standby mode if the patient requires continuous ventilation.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA