Operator’s Manual of Anesthesia System 10 - 15
Product Specifications Monitor Module
10.10 Monitor Module
10.10.1 AG Module
Flow Input Range: 20 L/min to 40 L/min
Maximum Vacuum: 69 kPa to 72 kPa (517.5 mmHg to 540 mmHg) with external
vacuum applied of 72 kPa (540 mmHg) and 40 L/min free
flow
Maximum Flow: 39 L/min to 40 L/min with external vacuum applied of 72
kPa (540 mmHg) and 40 L/min free flow
Minimum Flow: 20 L/min
Vacuum Gauge Accuracy: ± 5 % of full scale
TABLE 10-22 Suction device
Warm-up time: ISO accuracy mode: < 45 s
Full accuracy mode: <10 min
Sampling rate: Sampling rate:
Adult/Pediatric AG watertrap and sample line: 150/180/200 ml/min
Neonate AG watertrap and sample line: 100/110/120 ml/min
Accuracy: ± 10 ml/min or ± 10 % of the setting value, whichever is
greater
Watertrap emptying interval
1
Adult/Pediatric AG watertrap:
≥19h@150ml/min
≥18h@180ml/min
≥17h@200ml/min
Neonate AG watertrap:
≥24h@100ml/min
2
≥22h@110ml/min
≥20h@120ml/min
Gas: CO
2
, O
2
, N
2
O, and any of the five anesthetic agents: DES, ISO, ENF, SEV
and HAL.
Range: CO
2
: 0.0 to 30 % (0 to 30kPa, 0 to 226mmHg)
O
2
: 0 to 100 %
N
2
O : 0 to 100 %
DES : 0.0 to 30 %
SEV : 0.0 to 30 %
ENF : 0.0 to 30 %
ISO : 0.0 to 30 %
HAL : 0.0 to 30 %
TABLE 10-23 AG Module

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA