Setting Alarm Volume Alarms and Messages
8 - 6 Operator’s Manual of Anesthesia System
8.3 Setting Alarm Volume
The Alarms Volume settings adjust the audio level of all high, medium, and low priority sounding
alarms. The System Alert Volume settings adjust the audio level of all sounding pop-up prompts and
non-confirmed ventilation mode alerts.
To set the Alarm Volume:
1. On the Main Screen, select the soft key.
2. Adjust the volume on the pop-up screen.
Figure 8-4 Setting Alarm Volume
WARNING: Do not rely exclusively on the audible alarm system when using the
anesthesia system. Adjustment of alarm volume to a low level may
result in a hazard to the patient. Always keep the patient under close
surveillance.
NOTE: The auditory alarm signal A-weighted sound pressure level shall be no
less than 45dB and no more than 85dB.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA