Spirometry Screen Parameters and Factory Defaults
B - 2 Operator’s Manual of Anesthesia System
B.1 Spirometry Screen
B.2 Limits
ITEM DEFAULT
Loop Type P-V loop, F-V loop
Save Loop Baseline loop
Show Reference Off
Review Loops P-V loop
PARAMETERS DEFAULT
Alarms: Limits: PEAK High Adult: 50 cmH
2
O
Pediatric: 40 cmH
2
O
Neonate: 40 cmH
2
O
Alarms: Limits: PEAK Low Adult: 10 cmH
2
O
Pediatric: 8 cmH
2
O
Neonate: 8 cmH
2
O
Alarms: Limits: MV High Adult: 12 L/min
Pediatric: 6 L/min
Neonate: 6 L/min
Alarms: Limits: MV Low Adult: 1 L/min
Pediatric: 1 L/min
Neonate: 0.2 L/min
Alarms: Limits: Vte High 1000 ml
Alarms: Limits: Vte Low Off
Alarms: Limits: RR High Off
Alarms: Limits: RR Low Off
Alarms: Limits: EtCO
2
High Adult or Pediatric: 50 mmHg
Neonate: 45 mmHg
Alarms: Limits: EtCO
2
Low Adult or Pediatric: 25 mmHg
Neonate: 30 mmHg
Alarms: Limits: FiCO
2
High 4 mmHg
Alarms: Limits: Apnea Delay Time 30 sec
Alarms: Limits: CO
2
Apnea Delay Time 30 sec
Alarms: Limits: FiO
2
High Off
Alarms: Limits: FiO
2
Low 18 %
Alarms: Limits: EtO
2
High 88 %
Alarms: Limits: EtO
2
Low Off
Alarms: Limits: FiN
2
O High 53 %
Alarms: Limits: FiN
2
O Low 0 %
Alarms: Limits: EtN
2
O High 55 %
Alarms: Limits: EtN
2
O Low 0 %
Alarms: Limits: AGSS Alarms On
Alarms: Agents: EtHal High 3 %
Alarms: Agents: EtHal Low 0 %
Alarms: Agents: FiHal High 2 %

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals