Operator’s Manual of Anesthesia System 9 - 29
Maintenance Periodic Maintenance
Figure 9-38 Install the negative pressure suction device
NOTE: When install the filter to the suction tube, pay attention to keeping the
side printed with IN facing the liquid collection bottle.
NOTE: Avoid twisting or bending suction tubes during use.
9.7.4.5 Check Before the Next Use
After reassemble the components of the negative pressure suction device, check the negative
pressure suction device before use. See Section 5.15 "Inspect the Negative Pressure Suction Device".
9.8 Periodic Maintenance
WARNING: To avoid endangering the patient, do not perform test or maintenance
when the equipment is in use.
Visual inspection should be performed every 30 days to ensure timely replacement of worn or
damaged parts.
1. Power off the system.
2. Visually check the whole system.
3. Power on the system and follow the on-screen prompts to perform leak test and compliance
test. See Section 5.4.2 "Leak & Compliance Test".
Filter
Suction tube
Overfill protection connector

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals