System Information Title System Interface
4 - 6 Operator’s Manual of Anesthesia System
2. Set the network and ensure that the network access is normal.
a. Enter Standby mode.
b. Select the icon to open the [Setup] menu.
c. Select the [System] soft key, enter the system password and confirm the password.
d. Select the [Network] tab in the system menu.
e. Select [MD2] and set the [MD2] to (on) in the pop-up screen.
f. Set [Destination IP] and [Port].
g. Select the [Te st ] soft key.
h. Confirm that the test result is [Connected].
NOTE: Key information is defined by the eGateway. For specific
operations, see the eGateway Integrated Management and
Installation Guide.
4.2.2 Timer (Elapsed Timer/Countdown Timer)
Displays the elapsed timer, the countdown timer or both of the two timers. The timer is located in the
top left corner of the Main Screen. You can select the Timer icon to open the timer menu. (Figure 4-4)
Figure 4-4 Timer
Elapsed timer
Select the [Start] soft key to start the elapsed timer. Select the [Stop] soft key to stop the elapsed
timer. Select the [Reset] soft key to reset the timer.
Countdown timer
Set the remaining time for countdown and select the [Start] soft key to start the countdown.
Select the [Stop] soft key to stop the countdown timer. Select the [Reset] soft key to reset the timer.
The system will pop up a dialog box and beep when the countdown is over. In the dialog box, select
to turn off the system beep.
Timer icon

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals