Negative Pressure Suction Device Installation and Disassembly
3 - 6 Operator’s Manual of Anesthesia System
2. Insert the suction tube to the overfill protection connector. The negative pressure suction
device is now fully installed.
Figure 3-3 Install the negative pressure suction device
NOTE: When installing the filter to the suction tube, pay attention to keeping
the side printed with IN facing the liquid collection bottle.
NOTE: Avoid twisting or bending suction tubes during use.
3. To disassemble the negative pressure suction device, pull out the suction tube, take out the
liquid collection bottle and discard the filter. To replace the filter, please follow local regulations
to dispose of discarded filters.
3.5.1 Turn on Negative Pressure Suction Device
1. Assemble the negative pressure suction device.
2. Occlude the suction tube inlet at the patient end.
3. Turn on the negative pressure pipeline supply.
4. Set the selector switch to REG.
5. Adjust the negative pressure regulating knob to keep the pressure gauge reading smaller
than -40kPa.
3.5.2 Turn off Negative Pressure Suction Device
Set the selector switch to OFF to turn off the negative pressure suction device.
WARNING: Keep the negative pressure suction switch in the OFF state when the
negative pressure suction device is not in use.
Filter
Suction tube
Overfill protection connector
Liquid collection bottle

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals