Cleaning and Disinfection Maintenance
9 - 28 Operator’s Manual of Anesthesia System
Figure 9-37 Disassemble the negative pressure suction device
NOTE: Filters are disposable accessories. Please follow local regulations to
dispose of discarded filters.
9.7.4.2 Cleaning
1. Submerge or rinse the suction tube in an authorized cleaning agents (See Table 9-3 "Manual
Cleaning Agent" on Pages 9-9) for the required period as specified by the manufacturer of the
cleaning solution.
2. Rinse the suction tube with the clean water.
3. Completely dry the suction tube before use.
4. To disinfect the suction tube, proceed to 9.7.4.3 Disinfection. Otherwise, jump to 9.7.4.4
Reassembly.
9.7.4.3 Disinfection
NOTE: Before the disinfection, make sure that the suction tube has been
cleaned following 9.7.4.2 Cleaning.
1. Submerge or rinse the suction tube in an authorized disinfectors (See Table 9-4 "Manual
Disinfector" on Pages 9-10) for the required period as specified by the manufacturer of the
disinfecting solution.
2. Rinse the suction tube with the clean water.
3. Completely dry the suction tube before use.
9.7.4.4 Reassembly
NOTE: Before the reassembly, make sure that the assembly has been
reprocessed and dried.
NOTE: Reassemble the Negative Pressure Suction Device at the point of use
and at the designated cleaning area.
1. Place the liquid collection bottle to the bracket. Connect the suction tube and the filter
following the screen printed instructions.
2. Insert the suction tube to the overfill protection connector. The negative pressure suction
device is now fully installed.
Filter
Suction tube
Overfill protection connector

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals