EasyManuals Logo
Home>Mindray>Medical Equipment>A9

Mindray A9 Operating Instructions

Mindray A9
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #133 background imageLoading...
Page #133 background image
Operator’s Manual of Anesthesia System 6 - 11
Operations Set Ventilation
SIMV-VG means to deliver synchronized intermittent mandatory pressure control volume guarantee
ventilation to the patient. In the SIMV-VG mode, the ventilator waits for patient’s next inspiration
based on the specified time interval. The sensitivity depends on F-Trig (or P-Trig). If F-Trig (or P-Trig) is
reached within the trigger waiting time (called synchronous trigger window), the ventilator delivers
pressure controlled volume guarantee ventilation synchronously with the preset tidal volume and
inspiratory time. If the patient does not inspire within the trigger window, the ventilator delivers
pressure controlled volume guarantee ventilation to the patient at the end of trigger window.
Spontaneous breathing outside trigger window can acquire pressure support.
Set SIMV-VC, SIMV-PC or SIMV-VG Mode:
1. Select the SIMV-VC tab, SIMV-PC tab or SIMV-VG tab on the Main Screen.
2. Select the parameter soft key and assign an appropriate value on the pop-up screen.
3. Check that all SIMV-VC, SIMV-PC or SIMV-VG parameters are set appropriately.
4. Select the [Set Mode] soft key to confirm the settings.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals