Operator’s Manual of Anesthesia System 8 - 9
Alarms and Messages Setting Alarm Limits
Figure 8-8 Limits Tab
To set alarm limits:
1. On the Main Screen, select the [Alarms] soft key > [Limits] tab.
2. Select a parameter soft key.
NOTE: When the monitoring value on the Main Screen flashes due to an alarm, select
the flashing area to open the corresponding Alarm Limits setting menu.
3. Enter the desired parameter value using the keyboard on the screen. For each parameter,
the range of values is displayed above the keypad.
4. You can select [Load Alarm Defaults] to restore the default values and restore the upper and
lower limits of parameter alarms to the factory defaults.
5. Repeat Steps 3 and 4 for each parameter.
6. Select to save the changes (or select to discard the changes).
7. To save an alarm limit to a user profile:
Select the soft key > [System] soft key (system password required) > [Profiles] tab.
Select a profile, and select the [Create] soft key. Set the profile name on the pop-up screen.
After the profile is confirmed, the system will save the current profile as a user profile.
8. To load a user profile:
In the standby mode, select the [Current Profile: xxxx] soft key and select the desired
profile on the pop-up screen.
Select the [Manual] ventilation mode, and select the [Load Profile] soft key on the pop-
up screen, and then select the desired profile on the pop-up screen.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA