Operator’s Manual of Anesthesia System 1 - 5
Safety Safety Information
WARNING: Pins of connectors identified with the ESD warning symbol should not
be touched. Connections should not be made to these connectors
unless ESD precautionary procedures are used.
WARNING: Operation of the equipment below the minimum flow values may cause
inaccurate results.
WARNING: This equipment/system is intended for use by healthcare professionals
only. This equipment/system may cause radio interference or may
disrupt the operation of nearby equipment. It may be necessary to take
mitigation measures, such as reorienting or relocating the equipment,
or shielding the location it was placed.
WARNING: Ensure that an independent means of ventilation (e.g. a self-inflating
manual resuscitator with mask) is available whenever the equipment is
in use.
WARNING: The use of accessories with damaged packaging may cause
biocontamination or failure. The operator should check the integrity
of accessory packaging before use.
WARNING: Before using the anesthesia system after cleaning or disinfecting, power
on the system and follow the on-screen prompts to perform leak test and
compliance test. See (Pages 5-7) 5.4.2"Leak & Compliance Test".
WARNING: If the equipment is damaged in any way that compromises the safety of the
patient or user, discontinue use and attach a visible label indicating that the
equipment is unusable. Please contact Mindray Technical Support.
WARNING: Oxygen, when present in high concentrations, can significantly
increase the chance of fire or explosion. Oil and grease may be ignited
at the same time. Therefore, oil and grease should not be used where
oxygen enrichment may occur.
WARNING: Use of lubricants not recommended by Mindray may increase the danger of
fire or explosion. Please use lubricants as approved by Mindray.
WARNING: Low-pressure regulators and flow-meters are susceptible to high
pressure, and may burst if improperly maintained or disassembled
while under pressure. Changing or disassembling connectors should be
performed only by qualified personnel.
WARNING: Do not disassemble the low-pressure regulator, flow-metering device,
or connector while under pressure. Sudden release of pressure may
cause injury.
WARNING: Check the specifications of the Anesthetic Gas Scavenging System
(AGSS) and the specifications of the anesthesia system to ensure
compatibility and to prevent a mismatched processing system.
WARNING: Avoid connecting two or more hose assemblies in series as this may
cause a loss of pressure and flow.
WARNING: A hazard may exist due to the use of improper connectors. Ensure all
assemblies use the proper connectors.
WARNING: Avoid replacing a high-pressure flexible connection with one of lower
nominal inlet pressure.
WARNING: Reusing breathing circuits or reusable accessories that are not
disinfected may cause cross-contamination. Disinfect the breathing
circuits and reusable accessories before use.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals