Troubleshooting Your Analyzer
11-8
Recorder too hot
Recorder head too hot. 1. Stop using recorder and cool it for 10
minutes and then print again;
2. If the problem remains, shut down the
analyzer and restart it again, and then
print again;
1. If the problem remains, contact
Mindray customer service department.
Lyse out
Insufficient lyse or wrong lyse
volume setting.
1. Check if there is sufficient lyse left. If
so, access “Setup Settings
Reagents” and adjust the remaining
lyse volume as instructed by Chapter
5.3.1;
2. If not, change a new container of lyse
as instructed by Chapter 4.4.1.
Diluent expired
Expired diluent or wrong
expiration setting
1. Check if the diluent has expired. If so,
change a new container of diluent as
instructed by Chapter 4.4.1;
2. If not, reset the expiration date as
instructed in Chapter 5.3.1.
Rinse expired
Expired rinse or wrong
expiration setting
1. Check if the rinse has expired. If so,
change a new container of rinse as
instructed by Chapter 4.4.1;
2. If not, reset the expiration date as
instructed in Chapter 5.3.1.
Lyse expired
Expired lyse or wrong
expiration setting
1. Check if the lyse has expired. If so,
change a new container of lyse as
instructed by Chapter 4.4.1;
2. If not, reset the expiration date as
instructed in Chapter 5.3.1.
Vacuum filter
error
The air inside the vacuum
chamber is not extracted
within the given time.
1. Enter Service Self–test
Tubingto test the filter as instructed
in Chapter 10.5.1;
2. If the test result is normal, the error will
be removed;
3. If the problem remains, replace the
filter as instructed by Chapter 10.11;
4. If the problem still remains after a new
filter has been installed, contact
Mindray customer service department