8-6 Measurement
Table 5 B Performance
Probe D7-2s DE11-3s
C6-2Gs
6LB7s_C
6LB7s_L
C11-3s L16-4Hs
Expected
Lateral
Resolution
(mm)
≤
2(Depth
≤60)
≤
1(Depth
≤40)
≤
2(Depth
≤60)
≤
1(Depth
≤40)
≤
1(Depth
≤60)
≤
1(Depth
≤40)
≤
1(Depth
≤40)
Expected
Longitudinal
Resolution
(mm)
≤
1(Depth
≤80)
≤
1(Depth
≤60)
≤
1(Depth
≤80)
≤
1(Depth
≤60)
≤
1(Depth
≤80)
≤
0.5(Depth
≤60)
≤
0.5(Depth
≤60)
Expected Max
Detect Depth
(mm)
≥120 ≥80 ≥120 ≥80 ≥80 ≥80 ≥60
Probe C5-2s 6C2s 4CD4s P4-2s P7-3s P12-4s 7L4s
Expected
Lateral
Resolution
(mm)
≤1(Depth
≤80)
≤2(80<
Depth≤
130)
≤
1(Depth
≤40)
≤
2(Depth
≤60)
≤
3(Depth
≤80)
≤4(80<
Depth≤
160)
≤
2(Depth
≤60)
≤
1(Depth
≤40)
≤1(Depth
≤60)
Expected
Longitudinal
Resolution
(mm)
≤1(Depth
≤80)
≤2(80<
Depth≤
130)
≤
1(Depth
≤60)
≤
1(Depth
≤80)
≤
1(Depth
≤80)
≤2(80<
Depth≤
170)
≤
1(Depth
≤80)
≤
1(Depth
≤40)
≤
0.5(Depth
≤80)
Expected Max
Detect Depth
(mm)
≥180 ≥80 ≥120 ≥180 ≥80 ≥80 ≥80
Probe L7-3s L12-4s L14-6s L14-6Ns V10-4s V10-4Bs
Expected
Lateral
Resolution
(mm)
≤1(Depth
≤60)
≤1(Depth
≤60)
≤1(Depth
≤40)
≤1(Depth
≤40)
≤1(Depth
≤40)
≤1(Depth
≤40)
Expected
Longitudinal
Resolution
(mm)
≤1(Depth
≤80)
≤
0.5(Depth
≤80)
≤
0.5(Depth
≤60)
≤
0.5(Depth
≤60)
≤1(Depth
≤60)
≤1(Depth
≤60)
Expected Max
Detect Depth
(mm)
≥100 ≥80 ≥60 ≥60 ≥80 ≥80