EasyManuals Logo

Mindray M6 User Manual

Mindray M6
357 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #314 background imageLoading...
Page #314 background image
15-4 Batteries
15.6 Checking Battery Performance
The battery performance may be degraded over time, so you need to check the battery
performance periodically. The checking procedures are shown as follows:
1. Stop the ultrasound exam.
2. Connect the system to the AC power supply to charge the batteries until current
capacity is full.
3. Disconnect the system from the AC power supply to power the system by the
batteries until the system automatically shuts down.
4. The powering time of the batteries indicates the battery performance.
If the powering time of the battery is much shorter than that specified, you may replace the
battery or contact the service personnel.
NOTE:
The usage life of the battery is 300 charge / discharge cycles. If the battery is
improperly used, its life may be shortened.
15.7 Battery Disposal
You need to dispose of the battery when it is damages or severely degraded.
NOTE:
You should observe the local regulations when disposing of the battery.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M6 and is the answer not in the manual?

Mindray M6 Specifications

General IconGeneral
TypePortable Ultrasound System
ApplicationsAbdominal, Cardiac, Vascular, Small Parts, Musculoskeletal
Operating SystemWindows-based
PortsEthernet, USB
Supported ProbesConvex, Linear, Phased Array
Imaging ModesB-mode, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler

Related product manuals