ECG/RESP Monitoring
11-17
NOTE
z When monitoring a patient with a pacemaker, PACE must be turned ON.
Otherwise, the system will count the pacemaker pulse as QRS complex.
Do not completely depend on the alarms of heart rate. The patient with a
pacemaker must be nearly monitored.
z When PACE is turned ON, the system will neither detect the arrhythmia
relating to premature ventricular beats (including PVCs counting) nor
perform the ST analysis. When monitoring a patient without a
pacemaker, PACE should be turned OFF.
z The mode of PACE DETECT cannot be saved after the monitor is turned
off or be saved as the factory or the user default configuration. When the
monitor is started, restarted or when the factory or user default
configuration is selected, PACE DETECT returns to MODE1.
z The PACE tag can be printed in the real-time recording when PACE is
turned ON.
îš„ PACE LEAD
Set PACE LEAD as required.
To conduct the 5-lead ECG monitoring, you can select I,
II, III, aVR and aVL.
To conduct the 3-lead ECG monitoring, you can select I, II
and III.
NOTE
z You can switch the leads by setting LEAD TYPE in the ECG SETUP
menu. At the time of switching between 5 LEADS and 3 LEADS, PACE
LEAD in the ECG SETUP menu will be II.
îš„ NOTCH
Determines whether filter or not.
ON: The monitor protects the signals from the noise
generated by the power line.
OFF: No filtering.
During the real-time recording, the NOTCH on/off status
and the frequency will be recorded.