Setting the tuner frequency (Ftune)
Here you can set the tuner frequency and step width.
Setting the tuner frequency (Ftune)
You selected the desired view.
1.
Tap the Ftune button.
2.
Enter the required value and select a unit.
The settings is accepted immediately and the input window is closed.
Setting the step width (Fstep)
You selected the desired view.
1.
Long tap the Ftune button.
The context menu opens.
2.
Tap the Fstep button.
3.
Enter the step width and select a unit.
The settings are accepted immediately and the input window is closed.
4.
Press the Esc key to close the context menu.
Setting the channel / EMC bandwidth
(CBW / EBW)
Here you can do the following:
1.
Select the current filter type CBW or EBW (CISPR, MIL-STD-461-F).
2.
Set the channel bandwidth (CBW) or EMC bandwidth (EBW).
3.
Select Oversampling.
Normal and EMC filter types:
The channel bandwidth (CBW) filter type allows you to select a narrow frequency range
with the desired bandwidth from a wideband spectrum and to suppress the remaining
signal components.
Instead of the CBW filter two EMC filters according to the standards CISPR and MIL-
STD-461-F can be selected.
When selecting an EMC filter, following bandwidths are available:
CISPR: 200 Hz, 9 kHz, 120 kHz, 1 MHz
MIL: 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz
The available bandwidths also depend on the Fspan setting.
After selecting an EMC filter type, the label CBW is replaced everywhere in the
corresponding views by the label EBW.