SignalShark User Manual
90
Setting the resolution / EMC bandwidth (RBW / EBW)
You selected the desired view.
To set the resolution / EMC bandwidth:
1.
Tap the RBW / EBW button.
2.
Enter the required value and select a unit.
The setting is accepted immediately and the dropdown list is closed.
Settings in the context menu
You selected the desired view.
Long tap the RBW / EBW button.
The context menu opens.
Enabling/disabling the auto resolution bandwidth (Auto RBW /
Auto EBW)
Enable or disable automatic calculation of resolution bandwidth.
To set Auto RBW:
1.
Tap the Auto RBW button to switch function On / Off.
2.
Press the Esc key to close the context menu.
Setting the resolution bandwidth entry mode (RBW Entry / RBW Entry)
This button allows you to select the entry type Absolute or Relative.
To set RBW Entry:
1.
Tap the RBW Entry button to select the required type.
2.
Press the Esc key to close the context menu.
Setting the measurement time (Meas. Time)
Here you can set the measurement time. The measurement or detector time defines the time
range used by the internal detectors to calculate their values (+Pk, RMS, -Pk, ...).
To set the measurement time:
You selected the desired view.
1.
Tap the Meas. Time button.
2.
Enter the required value and select a unit.
The setting is accepted immediately and the dropdown list is closed.