4.1 Measuring with Constant Voltage or Current (ALC)
4-5 ZM2376
When you measure a high-dielectric type ceramic capacitor that has nonlinear characteristic, it is
recommended to use up to about 10µF at 1kHz and about 100µF at 120Hz. The measured value may
change due to signal distortion, when the voltage drop caused by the output impedance increases.
Figure 4-1 Maximum capacitance measurable at constant voltage (supplementary value)
Figure 4-2 Inductance range measurable at constant voltage (supplementary value)
100µ
1m
10m
100m
1
10µ 100µ 1m 10m 100m 1
電流 Arms
インダクタンス
H
Rd=100Ω
5V
Rd=
2
10mV
Rd=
2
5V
Rd=100Ω
10mV
Rd=6Ω
1.2V
上限
10µ
100µ
1m
10m
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1
試料端子電圧 Vrms
Cp max F
120Hz
1kHz
Rd: Output impedance
1kHz, Q =
Output impedance Rd = 6
D = 0.2, Sum of cable resistance and contact resistance = 0.5
Rd=6
can be used only at
1kHz, 40mA
Rd=25
can be used only when
Z1 (L159µH)
Rd=6 (measure range 100)
can be used only at 1kHz, 1.05V.
Measureable range
DUT terminal voltage Vrms
Upper limit
Lower limit
Indactance H
Current Arms