5.1 Error Message
5-4 ZM2376
5.1.2 Errors at Panel Operation
Main errors due to the panel operation are as listed below.
Error Message
Content and Cause
Corrective action
Settings conflict
Specified settings could not be made
due to the restriction by other settings.
Use within restricted conditions. Or,
change the settings that restrict others.
BIAS Over
DC bias voltage setting was limited so
that it does not exceed the maximum
output voltage (about 7Vpeak) because
measurement signal level is large.
Reduce the signal level when large DC
bias voltage is required.
LEVEL Over
Measurement signal level setting was
limited so that it does not exceed the
maximum output voltage (about
7Vpeak) because DC bias voltage is
output.
Reduce the DC bias voltage setting or
turn off the DC bias output when large
measurement signal level is required.
Memory error
Contents of setting/correction value
memory are lost.
Re-save the contents.
Since there are other error messages and warning messages not listed here, see the description of
respective operations.