5.2 When the Device Appears to be a Problem
5-10 ZM2376
Table 5-1 When the device appears to be a problem 4/4
Refreshing of display is slow.
Or, same measured value is
obtained repeatedly, and thus
the measured value does not
change.
No corrective action available.
When the measurement speed is fast, the device
thins out and displays the measured results so that
the measured values can be read appropriately.
The measurement itself is executed at the specified
speed.
Check the setting of averaging count, and set it to 1 if
the averaging is not necessary.
Check the trigger delay time, and set it to minimum
necessary value. When the trigger source is internal,
set it to initial value (8ms) or zero.
The measurement range is
switched by the range automatic
selection function.
The measurement range is not
determined due to a noise or
contact failure.
When many DUTs having almost same value are
measured, set the measurement range to HOLD.
The settling wait time with the automatic selection
enabled can be adjusted depending on the situation.