3.5 Basic Operations
3-24 ZM2376
Selection of measurement parameters
Selection of Primary Parameters
Press the _Pri_
key to display the primary parameter selection menu (see below), and select a
primary parameter with a numeric key. Also, the measurement parameters (primary and secondary)
can be selected automatically.
For details “3.5.5.1 Selection of Primary Parameters”
Primary parameter : C Auto:OFF Current setting
0)Auto OFF 1)ON 2)L 3)C 4)R 5)Z 6)Y 7)G Options
If either L, C, or R is selected, the equivalent circuit selection menu is displayed.
Selection of primary parameter equivalent circuit
For the L, C, and R, specify the equivalent circuit additionally.
Press the _CKT_
key to display the equivalent circuit selection menu (see below), and select
Parallel (suffix p) or Series (suffix s) with a numeric key. Automatic selection is also possible.
For details “3.5.5.2 Setting of Equivalent Circuit”
Equivalent CKT : Series Auto:OFF Current setting
0)Auto OFF 1)ON 2)Parallel 3)Series Options
Selection of secondary parameter
Press the _Sec_
key to display the secondary parameter selection menu (see below), and select a
secondary parameter with a numeric key. The options can be switched to those on the second page
with the _EXP_ | [NEXT] key.
For details “3.5.5.3 Setting of Secondary Parameters”
Secondary parameter : D Current setting
0)Q 1)D 2)θ 3)X 4)B 5)Rs >NEXT Options (first page)
6)Rp 7)G 8)Lp 9)Rdc <PREV Options (second page)
_CKT_ _Pri_ _Sec_